Olympus Pleuravideoscope, Model LTF-240
FDA Recall Z-1763-2020 · Posted None
Advertisement
Recalling FirmOlympus Corporation of the Americas
Recall StatusTerminated
Product CodeGCJ
Date Initiated2020-03-23
Date Terminated2021-11-30
Product Quantity265 total
Reason for Recall
There was no data to validate that the identified endo-therapy devices were compatible with certain pleuravideoscope models. Updated Operations Manuals are being provided.
Firm Action
An Urgent Medical Device Correction notification letter dated 3/23/20 was sent to customers.
Actions to be taken by the user:
Our records indicate that your facility has purchased one or more of the above-referenced
endoscopes. OLYMPUS requires that you take the following actions:
a) Inspect your inventory of operation manuals for the referenced Olympus models and
discard the existing ones.
b) Exchange them with the updated, enclosed Operation Manuals having an updated system
chart indicating compatible endoscope accessories.
c) Notify related personnel at your institution that they should stop the combination use of
the LTF-160 and LTF-240 and the subject Endo-Therapy accessories.
d) Access the Olympus recall portal to indicate that you have received this notification. Go
to https://olympusamerica.com/recall. Enter the recall number Recall-0066 and provide
your contact information as indicated in the portal.
e) If you have further distributed this product, identify your customers, forward them this
notification including the attachments, and appropriately document your notification
process.
Olympus regrets any inconvenience caused and fully appreciates your prompt cooperation in
addressing this situation. If you require additional information, please do not hesitate to contact
me directly at (484)896-5688 from Monday till Friday 8AM ET to 5PM ET or by e-mail at
[email protected].
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.