The PreludeSYNC Radial Compression Device is a sterile, single use disposable device used to assist in gaining and maintaining hemostasis of the radial and ulnar artery following catheterization procedures. It has a soft wristband with a secure hook and loop fastener. The band delivers adjustable compression of the puncture site with an inflatable bulb, and a check valve for easy inflation and deflation with a syringe. A clear curved backer plate provides optimal visualization of the puncture site and ease of placement.

FDA Recall Z-1766-2017 · Posted 2017-04-10

Recall StatusTerminated
Product CodeDXC
Date Initiated2017-03-03
Date Terminated2017-06-21
Product Quantity1,940 units total

Reason for Recall

Merit Medical Systems, Inc. announces a voluntary field action for the PreludeSYNC Radial Compression Device because particulate generated during the manufacturing process may enter the valve, preventing a complete seal of the inflation valve.

Firm Action

Merit Medical sent an Urgent Product Recall Notice on March 3, 2017, via letter for consignees to immediately stop using or distributing the affected lots and return them to Merit. For any questions, please contact Merit Customer Service at (801) 208-4381. For further questions regarding this recall, please call (801) 253-1600.

Root Cause

Process control

Distribution Pattern

Nationwide, Canadian and other foreign consignees. No VA/govt/military consignees.