KliniTray Platte Faxitron ,groB "R" sterile. Tissue resection margin examination board/large Model Number: 2001155-F The medical device KliniTrayTM is a system for the precise monitoring, recording and visualization of alignment during surgical procedures. It ensures the exact orientation and positioning of tissue to be removed in the patient's body, enabling improved surgical precision and patient safety.
FDA Recall Z-1769-2025 · Posted 2025-05-15
Advertisement
Recalling FirmKlinika Mdical Gmb
Recall StatusOpen, Classified
Product CodeFMH
Date Initiated2025-04-04
Product Quantity160 units
Reason for Recall
Potential for microbial contamination.
Firm Action
Single distributor was notified via email on 04/04/2025 with an "URGENT FIELD SAFETY NOTICE - PRODUCT RECALL" letter. The letter instructed the distributor to Identify and quarantine all units of the affected LOT numbers, do not distribute or use these products, treat the products as non-sterile (clean packed) and ensure they are not used on the sterile field, and inform all relevant departments or customers within your distribution chain about this FSCA. Additionally, the distributor was requested to complete and return the acknowledgment form by 04/16/2025, even if you no longer have affected stock.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the state of Massachusetts.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.