Open, Classified

Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat, Small, Cat. No. ARS1041102 3. TORNIER HRS Max Trial, Cat. No. ARS10252

FDA Recall Z-1769-2026 · Posted 2026-04-03

Recalling FirmTornier, Inc
Recall StatusOpen, Classified
Product CodeKWS
Date Initiated2026-03-05
Product Quantity233

Reason for Recall

Affected products were incorrectly labeled as Tornier HRS Max products. These parts are compatible with the Tornier HRS system only, and are incompatible with the Tornier HRS Max system.

Firm Action

On March 5, 2026, the firm notified affected customers of the product issue. Customers were instructed to locate affected product in their inventory. Relevant personnel should be informed that the affected product must not be used with Tornier HRS Max products, as they are incompatible. Product return is not necessary, but customers may contact the recalling firm to arrange for return if the product issue impacts their needs.

Root Cause

Labeling design

Distribution Pattern

US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KY, MA, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source