AquaC UNO H [120V, 60Hz, 1700VA, US-Version- a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-0970-1 Vivonic Part Number: G02000966-US
FDA Recall Z-1770-2022 · Posted 2022-09-19
Advertisement
Recalling FirmFresenius Medical Care Holdings, Inc.
Recall StatusOpen, Classified
Product CodeFIP
Date Initiated2022-06-21
Product Quantity6 units
Reason for Recall
Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
Firm Action
Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) issued Urgent Medical Deice Correction letter (FA-2022-20-C) on 06/21/22 via First Class Mail. Letter states reason for recall, health risk and action to take:
A Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) Water Systems Service
Specialist will be contacting you to schedule a correction of your portable Water Treatment
system. Only a Certified AquaC UNO H System Technician or FMCRTG Water Systems Service
Specialist may perform this service. If choosing to correct via your own Certified AquaC UNO H
System Technician, you will need to contact FMCRTG prior to performing the correction.
FMCRTG will provide the necessary instruction via a Change Notification.
For questions or concerns, please contact FMCRTG Technical Support Services at 888-553-1109.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.