Infusion System. EKOS EndoWave Infusion System. EKOS EndoWave Infusion System Drug Delivery Catheter with Ultrasound Core. Labeled as Sterile, Single Use Only. Connector clip is CIC Connector Clip. The infusion system is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytes, into the peripheral vasculature.
FDA Recall Z-1773-2008 · Posted 2008-08-20
Advertisement
Recalling FirmEKOS Corporation
Recall StatusTerminated
Product CodeKRA
Date Initiated2008-04-01
Date Terminated2011-05-13
Product Quantity218 units
Reason for Recall
Catheter disconnection. There is a potential for the catheter to disconnect from the Catheter Interface Cable (CIC) pod which connects to the Control Unit.
Firm Action
Firm representatives will visit and distribute connector clips and instructions for use to consignees beginning 04/01/2008 at no cost to the consignee.
Root Cause
Other
Distribution Pattern
Class II Recall - Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.