CardioCommand, Inc. Transesophageal Cardiac Pacing and Recording System, Model 7A, REF 0101 - Product Usage: Indications for Use: - Temporary left atrial pacing to treat bradycardia, hypotension or AV junctional rhythm - Acceleration of heart rate as an alternative to exercise or drugs during diagnostic cardiac studies such as echocardiography or radionuclide ventriculography. - Antitachycardia pacing for cardioversion of supraventricular tachycardia (SVT), including atrial flutter and re-entrant atrial or atrioventricular paroxysmal tachycardia. - Esophageal electrocardiography for differential diagnosis of complex atrial arrhythmia
FDA Recall Z-1773-2019 · Posted None
Recalling FirmCardiocommand Inc.
Recall StatusTerminated
Product CodeLPA
Date Initiated2019-05-10
Date Terminated2021-04-12
Product Quantity5 units
Reason for Recall
Possible corrosion of the battery contact(s).
Firm Action
The consignees were notified by letter on 05/10/2019. The notice requested the return of the unit.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of CT, VW, OH, IL, DE.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.