Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

See Luer Cap Set, MPC-130, set, administration, intravascular

FDA Recall Z-1774-2026 · Posted 2026-04-06

Recalling FirmMolded Products Inc
Recall StatusOpen, Classified
Product CodeFPA
Date Initiated2026-02-17
Product Quantity26900 units

Reason for Recall

Molded Products, Inc. is voluntarily conducting a medical device recall of the MPC-130 See Luer Cap Set based on a confirmed complaint of the threaded sleeve not being engaged and becoming un-attached.

Firm Action

Molded Products, Inc. issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 02/17/2026 via USPS First Class Mail. The notice explained the issue and requested the following: Please take the following actions: 1. Please inspect your inventory to determine if you have any of the affected lot. Quarantine the affected product and initiate a return to Molded Products, Inc. for credit. 2. Share this recall notification will all users of the product to ensure they are also aware of this recall. 3. Complete the attached Medical Device Recall Customer Response Form and return to the contact noted below whether or not you have any of the impacted material so that Molded Products may acknowledge your receipt of this notification and process your return and credit. For complaints, adverse events, or questions, contact Teresa Wooster, [email protected]; Ph# 800-435-8957 ext. 236 Monday Friday 8:00am-4:30pm (CT), Fax# 800-227-7935.

Root Cause

Process control

Distribution Pattern

US Nationwide distribution in the states of TN, TX, NV, IL, FL.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →