Open, Classified

Puritan Bennett 560 Ventilator, PB560, CFN 4096600

FDA Recall Z-1775-2022 · Posted 2022-09-22

Recalling FirmCovidien Llc
Recall StatusOpen, Classified
Product CodeQOS
Date Initiated2022-08-19
Product Quantity5 US

Reason for Recall

Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components.

Firm Action

Consignees were notified by letter delivered beginning August 19, 2022. Medtronic is advising that continued use of PB560 ventilators is appropriate when the actions listed in the recall letter are taken into consideration, unless this continued use is inconsistent with facility internal policies and procedures. All affected units require turbine replacement. Customers should contact Medtronic Service at 1-800-255-6774 to arrange for required service to replace the affected ventilator turbine component. If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.

Root Cause

Process control

Distribution Pattern

Domestic distribution to one customer in Washington, D.C. Foreign distribution worldwide.