Puritan Bennett 560 Ventilator, PB560, CFN 4096600
FDA Recall Z-1775-2022 · Posted 2022-09-22
Advertisement
Recalling FirmCovidien Llc
Recall StatusOpen, Classified
Product CodeQOS
Date Initiated2022-08-19
Product Quantity5 US
Reason for Recall
Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components.
Firm Action
Consignees were notified by letter delivered beginning August 19, 2022. Medtronic is advising that continued use of PB560 ventilators is appropriate when the actions listed in the recall letter are taken into consideration, unless this continued use is inconsistent with facility internal policies and procedures.
All affected units require turbine replacement. Customers should contact Medtronic Service at 1-800-255-6774 to arrange for required service to
replace the affected ventilator turbine component.
If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.
Root Cause
Process control
Distribution Pattern
Domestic distribution to one customer in Washington, D.C. Foreign distribution worldwide.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.