Terminated

Q-Logic Controller Software used on the following power wheelchairs: Q6000, Q6000XL, Q6000Z, R4000, Q600, Q600XL, Q610, Q1107, Q600E, and Dynamo 1107

FDA Recall Z-1776-2008 · Posted 2008-08-27

Recall StatusTerminated
Product CodeITI
Date Initiated2008-04-11
Date Terminated2009-11-13
Product Quantity4129

Reason for Recall

Software - Watch dog timer feature was disabled.

Firm Action

The recalling firm issued an Urgent Device Correction letter on 4/11/08. This letter instructs the dealers to contact the end users so that the units can be upgraded to enable the watch dog feature.

Root Cause

Software design

Distribution Pattern

Class II Recall - Worldwide Distribution --- including USA and Canada.