Q-Logic Controller Software used on the following power wheelchairs: Q6000, Q6000XL, Q6000Z, R4000, Q600, Q600XL, Q610, Q1107, Q600E, and Dynamo 1107
FDA Recall Z-1776-2008 · Posted 2008-08-27
Advertisement
Recalling FirmPride Mobility Products Corp
Recall StatusTerminated
Product CodeITI
Date Initiated2008-04-11
Date Terminated2009-11-13
Product Quantity4129
Reason for Recall
Software - Watch dog timer feature was disabled.
Firm Action
The recalling firm issued an Urgent Device Correction letter on 4/11/08. This letter instructs the dealers to contact the end users so that the units can be upgraded to enable the watch dog feature.
Root Cause
Software design
Distribution Pattern
Class II Recall - Worldwide Distribution --- including USA and Canada.