Grappler Suture Anchor, Titanium, 3.0 x 10mm, Sterile Kit, REF: P44-110-3010-SK, for use within the Grappler Suture Anchor System

FDA Recall Z-1776-2022 · Posted 2022-09-22

Recalling FirmParagon 28, Inc.
Recall StatusOpen, Classified
Product CodeMBI
Date Initiated2022-07-13
Product Quantity145

Reason for Recall

Suture anchor may break upon insertion, which could lead to potential debris in patient, and delays in surgery.

Firm Action

On 7/13/22, recall notices were emailed to distributors who were asked to do the following: 1) Review physical inventory of product within your control and at a hospital/health care facility within your region. Any affected product that is identified is to be discontinued, quarantined and not distributed any further. 2) Return affected devices to the recalling firm. 3) Complete and return the Recall Response form. For inquiries, contact the recalling firm at [email protected]

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution in the states of AZ, IN, LA, NJ, TX, WA, NY, MD, WI, OK, KY, MI, OH, CA, FL, KS, PA.