Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures.

FDA Recall Z-1776-2024 · Posted 2024-05-09

Recall StatusOpen, Classified
Product CodeHAW
Date Initiated2024-03-26
Product Quantity3286 applications.

Reason for Recall

Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.

Firm Action

On April 10, 2024, Medtronic Navigations, Inc. issued an Urgent: Medical Device Correction to affected consignees via UPS. In addition to informing consignees about the issue, Medtronic asked consignees to take the following actions: 1. Please review this information with all physician users. If you have any questions related to this issue, please contact Medtronic Technical Services for help at 1-888-826-5603 or email at [email protected]. 2. Please confirm via the enclosed confirmation form that this notification has been communicated within your facility with all physician users. Send the completed Customer Confirmation Form to Medtronic via email at [email protected]. a. Medtronic will work expeditiously with your facility to coordinate a software update at no charge to resolve this anomaly. 3. Pass this notice on to those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. 4. Adverse reactions or quality problems experienced with this product should be reported to FDA and Medtronic. Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail). Call FDA at (800) FDA-1088. Call Medtronic Technical Services at 1-888-826-5603.

Root Cause

Software design

Distribution Pattern

Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Republic Of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, Nicaragua, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.