Terminated

Boston Scientific Imager II Angiographic Catheter, M001316581 (5-pack outer package UPN), M001316580, single unit, inner package UPN)

FDA Recall Z-1781-2017 · Posted 2017-03-23

Recall StatusTerminated
Product CodeDQO
Date Initiated2016-12-13
Date Terminated2017-11-22
Product Quantity26 boxes of 5 (130 singles)

Reason for Recall

One lot of Imager II Angiographic Catheters is being recalled due to packaging mislabeling. The product may be labeled as Imager II Contra 2 curve catheters, when in fact the package contains Contra curve catheters.

Firm Action

Customers were sent an Urgent Medical Device Removal letter, dated 12/5/2016, via overnight mail delivery beginning 12/15/2016. The letter identified affected product, described the issue, and asked for product to be returned. Removal instructions were provided and a Reply Verification Tracking Form was asked to be returned. Questions can be directed to Nicole Pshon, Field Action Manager at 763-494-1133 or [email protected]

Root Cause

Employee error

Distribution Pattern

AZ, FL, MO, NJ, NY, OH, PA, and TX.