Recall StatusTerminated
Date Initiated2008-04-17
Date Terminated2009-12-02
Product Quantity180 units.
Reason for Recall
Breakage - Tips on the Pathfinder Bone Awl may bend or break prematurely. The firm has received internal non-conformances and complaints where the tips of the Awl were worn and damaged. Of the tip breaks, six have had material missing. During the same period, there were 15 complaints reported where the tip broke off at the weld after repeated use (1-3 years). The firm found several possible ca
Firm Action
Phone calls and emails were used to disseminate information about this recall. Firm notified consignees via telephone beginning on 04/17/08. Consignees were told that some problems were found through normal wear and tear as well as product mis-use and material weaknesses. The firm explained that they noticed from wear and tear from use as well as damage from use could cause the tip to become dull and flared causing breakage. The firm explained that they redesigned the awl and requested that consignees return outstanding inventory, which will be replaced with a new version of the awl.
Distribution Pattern
Class II Recall - Worldwide Distribution --- USA including states of AR, AZ, CA, CT, CO, FL, IL, MD, MI, MO, MT, NC, NJ, NV, NY, NM, OH, OK, PA, SD, TX and WI, and countries of Argentina, Australia, Belgium, Canada, China, Columbia, Cyprus, Germany, Greece, Italy, Japan, Mexico, Thailand, Turkey and UK/Ireland.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source