Cordis EMPIRA RX PTCA Dilatation Catheter Catalog # 85R30300S For cardiovascular use.

FDA Recall Z-1784-2014 · Posted 2014-06-11

Recalling FirmCordis Corporation
Recall StatusTerminated
Product CodeLOX
Date Initiated2013-02-05
Date Terminated2014-06-23
Product Quantity80 units

Reason for Recall

Dilatation catheters could exhibit radial versus axial tears should they burst during inflation.

Firm Action

An Urgent Medical Device Recall letter and Acknowledgement form was sent overnight to mulitple contacts in each account February 7, 2013. A representative will follow-up as necassary to facility obtaining signature, faxing the acknowledgment form to Cordis, collecting and returning units.

Root Cause

Process design

Distribution Pattern

Worldwide distribution: US states: AZ, FL, IL, LA, MA, NJ, OR, SC, and TX. Armenia, Austria, Belgium, Colombia, Czech Republic, France, Hungary, Iran, Israel, India, Italy, Kuwait, Latvia, Lebanon, Luxembourg, Mexico, Morocco, Poland, Portugal, Russia, Saudi Arabia, South Africa, Spain, Switzerland, and United Arab Emirates.