Open, Classified

Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system

FDA Recall Z-1784-2024 · Posted 2024-05-07

Recalling FirmNRT X-RAY A/S
Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2024-04-17
Product Quantity35

Reason for Recall

X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.

Firm Action

On 04/17/24, correction notices were emailed to consignees who were asked to do the following: 1) If you experience the described uncontrolled motion, stop using the device and contact your service provider to have your device repaired. 2) The correction notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. 3) Complete and return the response form to either [email protected] or [email protected] A firmware update will be made to all motor controllers in the system. The update will be performed by the Service Provider. Customers with recall-related questions can call 1-800-331-7327.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI.