STERIS Fine Traction Device, one unit per box Product Usage: The Fine Traction Device is used to provide traction and enable fine adjustments to traction and orientation of the leg during orthopedic procedures.

FDA Recall Z-1786-2013 · Posted 2013-07-22

Recalling FirmSteris Corporation
Recall StatusTerminated
Product CodeHST
Date Initiated2013-05-20
Date Terminated2013-08-28
Product Quantity16 units

Reason for Recall

A bearing used in the assembly of the Fine Traction Device did not perform successfully in pre-release reliability testing; a replacement bushing passed reliability testing and was substituted per a change order. The change order authorized use of the substitute bushing in assembly of the Fine Traction Accessory; however, inadvertently the new part was not implemented. During complaint evaluati

Firm Action

STERIS Account Managers notified affected customers via onsite visits. All consignees were notified as of 5/29/2013. Customers were instructed to return affected product for a replacement unit.

Root Cause

Component change control

Distribution Pattern

USA Nationwide Distribution in the states of CO, KY, LA, VA, and WI. The affected product was also distributed to Canada.