Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

FISTULA (ON/OFF) ALL IN ONE BAG. Medical convenience kit, Avid Medical Part Number: VAMK024-03

FDA Recall Z-1786-2021 · Posted 2021-06-04

Recalling FirmAvid Medical, Inc.
Recall StatusTerminated
Product CodeOEZ
Date Initiated2021-04-09
Date Terminated2022-07-18
Product Quantity160 kits

Reason for Recall

Medical convenience kits are being recalled due to a potential breach in the package integrity of a component, the BD ChloraPrep 3mL applicator. Under certain circumstances the applicator may grow Aspergillus penicillioides, which may result in a breach of the package integrity. A breach of package integrity may contaminate other components in AVID Medical kits.

Firm Action

A Recall notification letter titled, "URGENT MEDICAL DEVICE CORRECTION" was sent to consignees on 04/09/2021 via a email. End-users were notified on 4/13/21. The letter instructs the consignee to immediately examine all inventory locations and discontinue use of all Halyard or AVID branded Procedure Trays subject to the field action. Discard all affected Procedure Trays found in inventory. In addition, if the consignees has further distributed this product, the notice instructs the consignee to please identify their customers and notify them at once of this product correction and requirements. The notification to the consignee's customers may be enhanced by including a copy of this URGENT MEDICAL DEVICE CORRECTION Notice. The consignee is also asked to please complete the attached Field Action Response Form (page 3) immediately, even if they no longer have any of the recalled lot numbers left in inventory. The completed form is to be emailed to: [email protected]. If the facility has identified affected product in the consignee's inventory, the recall coordinator will reach back to their facility and provide the consignee with paperwork for documenting destruction of the recalled product.

Root Cause

Environmental control

Distribution Pattern

Distributed US nationwide to TX, AZ, IL, MS, LA, MO, and TN

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑