Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine Catalog Number: EV4115
FDA Recall Z-1786-2024 · Posted 2024-05-09
Recalling FirmRandox Laboratories Ltd.
Recall StatusOpen, Classified
Product CodeJJE
Date Initiated2024-04-05
Product Quantity40 units
Reason for Recall
Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.
Firm Action
Randox Laboratories Ltd UK notified the US Distributor Randox Laboratories via email on 4/05/ 24. End-Users notified by email on 4/9/24.
Letter (REC736) states reason for recall, health risk and action to take:
" Export support files for the last 10 samples run and share with Randox using the link
to be provided
" Discuss the contents of this notice with your Medical Director.
" Complete and return the response form 12187-QA to [email protected]
within five working days.
" Affected chip heater assembly replacement to be arranged by Randox Service
Engineer Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.