Jasper Vektor, Class II Correction Appliance, Part No. 610-527L used for correcting class 2 malocclusions by applying light forces on the teeth using a tube, arch wire and stops.

FDA Recall Z-1787-2018 · Posted None

Recalling FirmTP Orthodontics, Inc.
Recall StatusTerminated
Product CodeECO
Date Initiated2018-05-01
Date Terminated2020-04-15
Product Quantity0

Reason for Recall

The firm has received 106 complaints of the device breaking in the patient's mouth. Device breakage may cause injury to the patient.

Firm Action

On May 3, 2018, the firm sent an Urgent Medical Device Recall letter to customers. The letter advised customers that the firm was voluntarily recalling Generation 1 Jasper Vektor Class 2 Correction Appliances due to receiving 106 customer complaints that the device had broken in the patient's mouth during placement and usage. Customers were instructed to do the following: 1. Locate any of the affected lots identified in the table below and quarantine the product. 2. Return the Generation 1 Jasper Vektor for a replacement Jasper Vektor Generation 2 or refund, following the instructions below: Contact TP Orthodontics, Inc. at 800-338-8856 or [email protected] for a Return Goods Authorization and pre-paid shipping label. Alternately, use the carrier of your choice with available tracking and send to: TP Orthodontics, Inc. ATTN: Recall Coordinator 100 Center Plaza La Porte, IN 46350 3. Email, mail, or fax the attached response letter to TP Orthodontics, Inc. In addition to sending the letter back, you may call us with any questions or concerns. EMAIL: [email protected] FAX: 219-324-3029 MAIL: TP Orthodontics, Inc. ATTN: Recall Coordinator 100 Center Plaza La Porte, IN 46350 4. If you have any further questions or concerns after reviewing this notice, please contact TPO Customer Care Team 800-388-8856 or [email protected]

Root Cause

Device Design

Distribution Pattern

Distributed to accounts in 20 states: AL, CA, CO, FL, HI, ID, IL, MA, ME, MI, NJ, NV, NY, OH, OR, PA, SC, TX, UT, and VA. Foreign distribution to United Arab Emirates, Australia, Belgium, Brazil, Canada, Germany, Spain, Finland, France, Great Britain, Hungary, Ireland, Italy, Japan, Mexico, Netherlands, New Zealand, Portugal, Singapore, and South Africa.