GE Healthcare Revolution XR/d; GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model numbers: 2259988-2 & 2351505. The Revolution XR/d Digital Radiographic Imaging System is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in all general-purpose diagnostic procedures. This device is not intended for mammographic applications.
FDA Recall Z-1788-2012 · Posted 2012-06-08
Recalling FirmGeneral Electric Med Systems LLC
Recall StatusTerminated
Product CodeMQB
Date Initiated2006-05-31
Date Terminated2012-07-31
Product Quantity3,796 units
Reason for Recall
An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath
Firm Action
GE Healthcare sent an "URGENT SAFETY INFORMATION" letter dated May 31, 2006 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included new operator instructions and warning labels.
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution.