Nimbus 4 and Nimbus Professional Alternating-pressure bed mattress overlay system
FDA Recall Z-1788-2021 · Posted None
Advertisement
Recalling FirmGetinge (Suzhou) Co Ltd
Recall StatusTerminated
Product CodeFNM
Date Initiated2021-04-15
Date Terminated2023-11-13
Product Quantity13042 units
Reason for Recall
There is a potential for abnormal inflation of the mattress.
Firm Action
Customers identified as owners of the affected mattresses will be notified of the identified product risk and the Field Safety Corrective Action through the Field Safety Notice.
The Field Safety Notice will be accompanied by Customer Response Form.
The Field Safety Notice will be accompanied by a video recording. The video demonstrates how to remove the over-inflation risk immediately through puncture of two grommet membranes in the mattress that can be performed on-site.
Throughout technical evaluation has confirmed that the puncture of grommet membranes eliminates the patient fall risk and does not introduce any new harm.
The customer will be asked to acknowledge the FSN, complete and sign an enclosed Customer Response Form and return it to the local Arjo office (address indicated in the FSN) in order to confirm that the advised activity has been completed following the provided instruction.
Root Cause
Under Investigation by firm
Distribution Pattern
USA
Official Source