Columbus Tibia Plateau (component of the Columbus REVISION Knee System) Product Usage - The Columbus Revision Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplastics or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture

FDA Recall Z-1793-2012 · Posted None

Recall StatusTerminated
Product CodeJWH
Date Initiated2012-02-21
Date Terminated2014-01-15
Product Quantity8

Reason for Recall

During an internal quality inspection, the tibia plateau was found to be out of specification. There is a potential for misalignment between the tibia plateau and attachment of the extension stem.

Firm Action

Aesculap Implant Systems (AIS) initiated their recall of this product by first telephoning consignees and then on February 21, 2012 by sending an IMPORTANT RECALL NOTIFICATION letter dated February 21, 2012 to all consignees. The letter identified the affected product, problem and actions to be taken. The notification told consignees not to use the affected product and to return any stock to AIS (USA) promptly. For questions contact Scott Stephan at (610) 984-9239 or Christian Gabriel at (610) 984-9300.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution including PA and TX.