MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
FDA Recall Z-1794-2026 · Posted 2026-04-08
Advertisement
Recalling FirmBeckman Coulter, Inc.
Recall StatusOpen, Classified
Product CodeLTT
Date Initiated2025-09-08
Product Quantity3 boxes
Reason for Recall
Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.
Firm Action
On 09/08/2025, the firm sent via mail/email an "URGENT FIELD SAFETY NOTICE" to customers informing them that as a result of an internal investigation Beckman Coulter confirmed that Ciprofloxacin (Cp) antimicrobial agent was incorrectly manufactured on MicroScan Neg Multidrug Resistant MIC 1 Lot 2026-05-08.
Customers are instructed to:
"Review Ciprofloxacin results generated from Lot 2026-05-08 to identify any abnormal or unexpected trends in both susceptible and resistance patterns, or unexpected change in MIC trends, as compared to historical results for that same species (e.g. a sudden, unexplained increase in susceptible or % susceptible results for a species after implementation of Lot 2026-05-08 versus the previous susceptible or % susceptible results for the same species prior to using Lot 2026-05-08).
"Discontinue use of this lot and discard per your laboratory guidelines
"The laboratory should retain the inventory of other MicroScan Neg Multidrug Resistant MIC 1 lots as they are not impacted by this issue.
Note-Beckman Coulter is no longer distributing MicroScan Neg Multidrug Resistant MIC 1 Lot 2026-05-08
For questions - contact Customer Support Center from website: http://www.beckmancoulter.com
Root Cause
Under Investigation by firm
Distribution Pattern
International distribution to the country of Poland.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.