Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging.

FDA Recall Z-1795-2013 · Posted 2013-07-24

Recall StatusTerminated
Product CodeMQB
Date Initiated2013-06-05
Date Terminated2015-02-24
Product Quantity401

Reason for Recall

Siemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and releasing the table top, it is possible that the fingers of the patient or operator are between the table top and the detector tray and if the table top is moved, the fingers can be squee

Firm Action

On 6/5/13, consignees were notified by letter of a "Field Safety Notice" and received instructions about how to avoid the potential risk of this issue while using the device. Siemens also rolled out a modification of AXIOM Aristos MX that will avoid these potential injuries in the future.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution.