Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Medtronic INTREPID Spinal System SPACER, 32X23, 12 DEG, Size: 10MM, Part Number: REF 7961210; 12MM, REF 7961212; 14MM, REF 7961214; 16MM, REF 7961216; 18MM, REF 7961218; 20MM, REF 7961220; Rx only, STERILE R The INTREPID Spinal System is indicated for use with autogenous bone graft in patients with degenerative disc disease.

FDA Recall Z-1797-2009 · Posted 2009-07-24

Recall StatusTerminated
Product CodeMAX
Date Initiated2009-02-06
Date Terminated2010-04-19
Product Quantity506 units

Reason for Recall

The firm received complaints regarding the use of the product in patients with poor bone quality, translational instability, difficulty implanting the cover plate and case of cover plate detachment.

Firm Action

Medtronic Sofamor Danek USA, Inc. instructed Sales Representative on starting February 6, 2009 by phone to immediately return product from the field. The firm further issued a "Subject: Medtronic Intrepid Intervertebral Body Fusion Device Recall" notice dated April 2009. The recall notices were mailed directly to Risk Managers and Surgeons, with Recall Questionnaires and Surgeon Confirmation forms included respectively. For further questions, contact your Medtronic Sales Representative or the Medtronic Global Quality Department at 1-800-876-3133 extension 6333.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution -- United States, Australia, Germany, The Netherlands, and South Africa.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑