Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

MicroVue BAP EIA, Model 8012, provides a quantitative measure of bone-specific alkaline phosphatase (BAP) activity in serum.

FDA Recall Z-1797-2018 · Posted None

Recall StatusTerminated
Product CodeCIN
Date Initiated2016-07-14
Date Terminated2020-08-28
Product Quantity1004

Reason for Recall

There is a possibility of low volume and/or leaking standard bottles.

Firm Action

Customers were contacted via telephone and a letter dated 7/14/16. The letter instructed customers to review their inventory for the specified lot number, check for any low-volume or leaking standards, and contact the recalling firm if necessary.

Root Cause

Process control

Distribution Pattern

The products were distributed to the following US states: CA, CT, GA, IL, KY, MA, MD, MO, PA, TX, WA, and WI. The products were distributed to the following foreign countries: Australia, Brazil, Canada, China, Costa Rica, India, Japan, Mexico, The Philippines, Russia, South Korea, South Africa, Switzerland, and Uruguay.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑