MoXy Laser Fiber, Material Number 0010-2400 - Product Usage: The fiber can be used for the surgical incision/excision, vaporization, ablation, hemostasis and coagulation of soft tissue.
FDA Recall Z-1798-2020 · Posted None
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusTerminated
Product CodeGEX
Date Initiated2020-03-30
Date Terminated2023-04-05
Product Quantity146,633
Reason for Recall
To reinforce existing instructions within the Directions for Use (DFU) and provide further guidance which will be added to the DFU that may reduce the likelihood of metal cap and fiber tip break temperature related complaints.
Firm Action
An Urgent Medical Device Product Advisory - Immediate Action Required dated 3/30/20 was sent to customers.
Existing MoXy" laser fiber Direction for Use to be Reinforced
1- Ensure the distance from the fiber to the tissue is approximately 2mm (1 to 3 mm).
Reference: Table 2. Troubleshooting and Procedural Considerations in product DFU.
2- Do not bury the tip of the fiber into the tissue. Reduced irrigation flow may result in damage
to the fiber. Reference: Caution in page 3 of product DFU.
3- Failure to position the saline solution at least 106cm or not opening the flow valve will
decrease the amount of fluid across the fiber end. Decreased fluid flow may limit the lasing
time of the fiber. Reference: Caution in page 3 of product DFU.
Section 8.3 of the MoXyTM Laser Fiber DFU has been updated to recommend an increased
irrigation flow to increase the liquid cooling effect. The additions to the current DFU are
highlighted in blue:
8.3 Performing the Procedure
1. Ensure user, patient, and personnel in the room are wearing appropriate protective eyewear
before using the laser.
WARNING: Do not press the footswitch without having the appropriate protective eyewear.
Pressing the footswitch may activate the laser beam. Activation of the laser beam when not
wearing protective eyewear may result in eye injury to the user, personnel in the room, or the
patient.
2. Connect the tubing from the sterile saline for irrigation solution to the fiber luer lock connector.
Position the saline solution at least 106 cm higher than the level of the endoscope/cystoscope.
Increased irrigation flow by means of a saline pressure bag (set to 250 mmHg 300 mmHg) will
further increase liquid cooling effect and may reduce fiber tip damage.
Please complete and return the acknowledgement form to indicate receipt of this letter.
Boston Scientific Quality Systems
763-494-1133
[email protected]
Root Cause
Use error
Distribution Pattern
Worldwide distribution - US Nationwide and Global distribution.