JELCO HYPODERMIC NEEDLE-PRO FIXED NEEDLE INSULIN SYRINGE: a) 1.0mL 28G X 1/2" (0.33mm x 13mm), REF 4428-1 b) 1.0mL 29G X 1/2" (0.33mm x 13mm), REF 4429-1

FDA Recall Z-1798-2021 · Posted 2021-06-10

Recall StatusTerminated
Product CodeMEG
Date Initiated2021-05-17
Date Terminated2023-10-11
Product Quantity1,432,200 units

Reason for Recall

Specific models and lot numbers of the Jelco Hypodermic Needle-Pro Fixed Needle Insulin Syringe 29Gx1/2 odd number line graduation markings on the syringe barrel were skewed approximately 20 degrees upward.

Firm Action

The firm initiated the recall on May 17, 2021 by email. The firm is seeking the removal of the product down to the consumer level. For assistance ease contact Novasyte using the Email: [email protected] The firm later realized they failed to include lot #40140906 for model #4429-1 and issued a second letter which was dated 5/27/2021 to customers involved in the first notification where this lot was excluded in the letter, as well as to new consignees who received the affected lot.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of UK, Spain.