Completed

Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA Test; Material Number 425-2260; Lot number 18200A - Product Usage: The assay is an aid in the diagnosis of celiac disease (gluten-sensitive enteropathy) and should be used in conjunction with other serological tests and clinical findings.

FDA Recall Z-1800-2020 · Posted None

Recalling FirmBio-Rad Labs
Recall StatusCompleted
Product CodeMVM
Date Initiated2019-04-03
Product Quantity376 units

Reason for Recall

An elevated lot-to-lot bias was observed for Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA when compared to a previous lot of product. The observed elevated bias between these two lots is largely due to samples with higher levels of antibodies. With the cut-off for the assay being 15 U/mL, the clinical classifications for most samples would remain unchanged.

Firm Action

On April 3, 2019, a Field Action Letter with response form was sent to U.S. Customers via Fed Ex courier while subsidiaries were notified via email. Customers were instructed to immediately discontinue further use of, and dispose of, any unused product according to local waste management procedures. Customers were requested to complete the attached customer response form as an acknowledgement with information including: account name, customer account number, contact name and title, address, and telephone number. Customers were instructed to check off the statement acknowledging awareness of the field action and that they had proceeded according to instructions issued by Bio-Rad. Completed forms were to be returned to [email protected]. Customers were instructed to contact their local Bio-Rad Technical Support with any questions regarding the Field Action Notification.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide distribution including the states of AZ, NY, LA, FL, CA, TX, IA, MS, UT, and IL.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →