EZPass Suture Retriever, Nitinol, Single Pack, Part number 10007380. Orthopedic Manual Surgical Instrument, Sterile, single use arthroscopic instrument utilized in conjunction with a nitinol wire to aid in passing suture through soft tissue.
FDA Recall Z-1801-2014 · Posted 2014-06-16
Advertisement
Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeLXH
Date Initiated2014-04-15
Date Terminated2015-07-07
Product Quantity100 units
Reason for Recall
The nitinol wire or monofilament may buckle within the EZPass Suture Passer and become kinked making the wire/monofilament difficult to use.
Firm Action
On 2/28/14, customers received a letter stating these devices have been recalled following an investigation which identified that the nitinol wire or monofilament may buckle within the EzPass Suture Passer and become kinked making the wire/monofilament difficult to use. The letter also stated, there is no adverse health outcome expected for the patient and this action requires the immediate location and discontinued use of the product and its return to Biomet.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution.