RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.

FDA Recall Z-1806-2014 · Posted 2014-06-17

Recall StatusTerminated
Product CodeMUJ
Date Initiated2014-04-25
Date Terminated2017-06-26
Product Quantity25 sites/245 units (US), 25 sites/124 units (foreign)

Reason for Recall

This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.

Firm Action

On April 25, 2014, RaySearch emailed all consignees of the affected product a copy of the recall notice as well as the response forms. Back up notification methods include phone calls and site visits.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution-USA (nationwide) including the states of AZ, CA, FL, IL, MA, MD, MO, MS, NJ, NV, NY, OH, PA, TN, VA, and WA, and the countries of Australia, Belgium, Germany, France, Italy, New Zealand, Republic of Korea, Switzerland, Spain, and United Kingdom.