META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690

FDA Recall Z-1806-2025 · Posted 2025-05-19

Recalling FirmSmith & Nephew, Inc.
Recall StatusOpen, Classified
Product CodeHSB
Date Initiated2025-03-18
Product Quantity75

Reason for Recall

Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.

Firm Action

On 3/18/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) Inspect your inventory and locate any affected devices and quarantine them. 2) If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. 3) Complete and return the notice response section via email to [email protected] If you have any questions email the firm at the email address above.

Root Cause

Process control

Distribution Pattern

Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA.