Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Swift Mobil shower chair, Swift Mobil 160 shower chair, and Swift Mobil Tilt shower chair. Height adjustable four wheeled shower chairs for persons of various weights.

FDA Recall Z-1807-2013 · Posted 2013-07-25

Recalling FirmEtac Supply Center Ab
Recall StatusTerminated
Product CodeINN
Date Initiated2013-05-23
Date Terminated2014-10-24
Product Quantity7200 internationally

Reason for Recall

Etac is voluntarily conducting a field correction of the Etac Swift Mobil shower chair. If the Swift Mobil shower chair loses a caster and thus becomes instantly unstable, there is potiential risk for harm or injury.

Firm Action

Etac sent an "URGENT FIELD CORRECTION" letter dated May 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.

Root Cause

Component change control

Distribution Pattern

International distribution: Austria, Australia, Belgium, Brazil, Canada, Czech Republic, Denmark, Esonia, Finland, France, Great Britain, Germany, Greece, Hong Kong, Ireland, Isreal, Iceland, Italy, Kingdom of Bahrain, Kuwait, Lithuania, Netherlands, Norge, New Zealand, Spain, Switzerland, Sweden, and United Arab Emirates.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑