The product is a handheld ultraviolet-C germicidal wand and contains a combination of five UV-A and UV-C LED lights. The product comes in two configurations: UV-C x 1; UV-A x 4, or UV-C x 2; UV-A x 3. The product has two peak emission wavelengths at 274 nm and 397 nm.
FDA Recall Z-1808-2026 · Posted 2026-04-17
Advertisement
Recalling FirmUvlizer c/o RAIS INTERNATIONAL LLC
Recall StatusOpen, Classified
Product CodeRHP
Date Initiated2026-03-06
Product Quantity334
Reason for Recall
In some foreseeable use conditions, the products can expose nearby persons to
UVC radiation at levels significantly above limits recommended by international safety guidelines
for skin and eye exposure.
Firm Action
Notifications to purchasers and subsequent transferees will be transmitted through multiple
communication channels: Certified Mail, Email Notifications, and Public Website Notice, describing the defect, safety considerations, and
instructions to discontinue use of the Product.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
U.S.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.