Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system Interventional fluoroscopic x-ray system System code: 722079

FDA Recall Z-1809-2020 · Posted 2020-04-29

Recall StatusTerminated
Product CodeOWB
Date Initiated2020-04-03
Date Terminated2023-02-01
Product Quantity22 units (6 US and 16 OUS); Expanded 3units; Total: 25

Reason for Recall

When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.

Firm Action

Philips Healthcare issued Urgent Medical Device Correction letter via Field Action 2020-IGTBST-003 issued on April 3, 2020. Letter states reason for recall, health risk an action to take: To avoid this hazardous situation, the users should: BY CUSTOMER/ USER " Use XperGuide live guidance with the L-arm positioned in the XperCT scan position. " Move the L-arm back to the initial XperCT scan position if the system displays the warning Live 3D roadmap not possible. Please move stand to the : [acquisition position] position. At any stage during needle guidance, the user can perform a verification run to verify if the needle position is correct. Please ensure that all staff working with the XperGuide application are informed of the content of this letter and place a copy of this letter as an addendum to the Instructions for Use. Questions contact your local Philips representative. Technical Support Line: 1-800-722-9377. Philips will resolve the problem for all affected systems by installing a new Interventional Workspot software release. This action will start in April 2020. Expanded Recall 12-22-20: The two systems for the Netherland have already been corrected through the implementation of software update. For the system in Germany, the Customer has been notified through the Field Safety Notice.

Root Cause

Software design

Distribution Pattern

Worldwide distribution. US states of FL, ID, IL, IN, MA, and OH, Canada, Australia, Austria, Belgium, Germany, and Netherlands New Zealand, South Africa, Switzerland, Thailand