Terminated

CyberKnife Treatment Delivery System, a subsystem of the CyberKnife Robotic Radiosurgery System, manufactured by Accuray Inc., Sunnyvale, CA. Indicated for treatment for planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

FDA Recall Z-1810-2009 · Posted 2009-09-14

Recalling FirmAccuray Inc
Recall StatusTerminated
Product CodeIYE
Date Initiated2009-06-22
Date Terminated2011-01-05
Product Quantity61 units

Reason for Recall

Targeting accuracy out of specification, Error alert does not render the system down, which may result in mistreatment in the wrong area.

Firm Action

Accuray, Inc. issued an "Urgent Device Correction" notice dated June 22, 2009. The consignees were informed of the defect including methods for detecting the error. The firm is in the process of developing a software patch to address the issue. For further information, contact Accuray, Inc. Customer Support at +1-877-668-8667 (US) or +1-408-716-4700 (outside US) or [email protected].

Root Cause

Software design

Distribution Pattern

Worldwide Distribution -- US including states of MA, FL, AL, TN, DE, CA, NJ, OK, CO, NC, IL, PA, MD, DC, MS, CN, NV, RI, SC, UT, VA, MI, NE, MO, TX and AL and countries of Canada, Russia, UK, Ukraine, Saudi Arabia, Switzerland, Turkey, Taiwan, India, China, Korea and Thailand.