Alaris PC unit model 8015, Infusion pump. software version 9.17 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The software is embedded in the Alaris PC unit.
FDA Recall Z-1811-2015 · Posted 2015-06-16
Advertisement
Recalling FirmCareFusion 303, Inc.
Recall StatusTerminated
Product CodeFRN
Date Initiated2015-05-13
Date Terminated2016-05-20
Product Quantity63 units
Reason for Recall
An issue with the cancel functionality used during atypical infusion programming to cancel user inputted values. An infusion may start that is greater than or less than the hospital established limits for the specific medication.
Firm Action
CareFusion sent an Urgent Medical Device Recall Notification letter dated May 13, 2015 to customers. The letter identified the affected product, problem and the recommended actions to be taken. Customers are informed that CareFusion has released a software upgrade to address this issue. CareFusion will contact all affected customers within 60 days to initiate the scheduling process for the software upgrade installation. Customers with recall related questions are instructed to contact CareFusion Support Center at (888) 562-6018, 7am-4pm PT, [email protected]. Customers with adverse event report questions are instructed to contact customer advocacy at (888) 812-3266, 24 hrs/day, 7 days a week, [email protected]. Customers with technical questions regarding the Alaris System are instructed to contact Technical Support at (888) 812-3229, 6am-5pm PT, Mon-Fri, [email protected].
Root Cause
Software design
Distribution Pattern
US Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.