Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

BD Affirm VPIII Microbial Identification Tests, packaged in kits, 120 test\kit, Catalog # 446257 and 24 test\kit, Catalog # 446252; kits are labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle Sparks, MD 21152 USA, 800-638-8663, www.bd.com/ds*** The Affirm VPIII Microbial Identification Test is a DNA probe test intended for use in the detection and identification of Candida species,

FDA Recall Z-1812-2013 · Posted 2013-07-26

Recall StatusTerminated
Product CodeMLA
Date Initiated2013-06-13
Date Terminated2013-09-28
Product Quantity177 kits

Reason for Recall

In vitro diagnostic test kit may exhibit invalid internal negative control results.

Firm Action

Becton Dickinson Diagnostic Systems notified distributors by fax on June 13, 2013, and advised consignees to discontinue use and discard affected product for credit. Distributors were requested to return their distribution list to BD for direct notification by the recaller. End users were contacted by letter on June 13, 2013 and the notification addressed to Microbiology Lab Director/Risk Managers advised that a small number of tests associated with the referenced lots of product may exhibit invalid internal negative control results. If encountered, the issue would result in an invalid test, as described in the package insert. End users were requested to discontinue use of these lot numbers, discard any remaining packages for replacement and return the enclosed response form. Customers with questions were instructed to contact BD Technical Services Department at 1-800-638-8663. For assistance regarding replacements customers were instructed to contact BD Customer Service at 1-800-675-0908. For questions regarding this recall call 410-316-4054.

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide Distribution

Official Source

View FDA Recall →