Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Tosoh ST AIA-PACK HbA1c Pretreatment Solution, HbA1c Pretreatment Solution, PN 025718 The device is intended for In Vitro Diagnostic Use Only for the pretreatment of the patient samples or control set for the HbA1c assay

FDA Recall Z-1812-2018 · Posted None

Recalling FirmTosoh Bioscience Inc
Recall StatusTerminated
Product CodeN/A
Date Initiated2018-03-05
Date Terminated2020-06-16
Product Quantity2258

Reason for Recall

The assay can potentially generate erroneously elevated or erroneously decreased HbA1c patient results.

Firm Action

On March 5, 2018, the firm notified all customers who have received product since 2011 of the discontinuation of this product from the market. Customers were instructed to immediately discontinue use of all lot numbers of recalled test and components, remove all recalled materials from inventory, and destroy per the customer's local procedures. Customers were also instructed to notify and forward the information to physicians who have received test results provided by recalled products over the past 6 months to determine if follow up testing is required. Customers were advised that the recalling firm is permanently removing all ST AIA-PACK HbA1c test and components from the market, and were instructed to identify an alternative test option for HbA1c. Customers were asked to complete and return the Recall Acknowledgement Form to the recalling firm. Customers may contact Tosoh Technical Support at 1(800)-248-6764, 8AM-5PM PST (voice mail after hours), seven days a week.

Root Cause

Nonconforming Material/Component

Distribution Pattern

USA (nationwide) Distribution including PR and to the states of : AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, MI, MO, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑