ARTIS Pheno VE30A and VE40A, Model 10849000

FDA Recall Z-1814-2026 · Posted 2026-04-16

Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2026-03-12
Product Quantity73

Reason for Recall

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

Firm Action

Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers dated March 9, 2026.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

U.S. and O.U.S.