ARTIS Pheno VE30A and VE40A, Model 10849000
FDA Recall Z-1814-2026 · Posted 2026-04-16
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2026-03-12
Product Quantity73
Reason for Recall
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
Firm Action
Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to
all affected customers dated March 9, 2026.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
U.S. and O.U.S.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.