Hospira Phoenix Infusion System with MedNet Software, Symbiq Two-Channel Infusion System, Software versions v2.1 & v3.0, Manufactured by Hospira Inc., Morgan Hill, CA

FDA Recall Z-1816-2009 · Posted 2009-08-11

Recalling FirmHospira Inc
Recall StatusTerminated
Product Codefrn
Date Initiated2009-03-31
Date Terminated2010-12-18

Reason for Recall

Potential delay/underinfusion of critical therapy-- Devices experience increased frequency of alarms that result in temporary pump stoppage.

Firm Action

Hospira, Inc. issued an "Urgent: Device Field Correction" dated March 31, 2009 via Federal Express to Healthcare Professionals and Valued Hospira Customers. Consignees were informed of the affected devices and will be contacted by a Hospira representative to arrange necessary device upgrades. For further questions, contact Hospira Global Product Safety and Complaints at 1-800-441-4100.

Root Cause

Software design

Distribution Pattern

Product was distributed to 59 consignees to all 50 States.