Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599 Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation.
FDA Recall Z-1816-2013 · Posted 2013-07-29
Advertisement
Recalling FirmSynthes USA HQ, Inc.
Recall StatusOpen, Classified
Product CodeKTT
Date Initiated2012-10-12
Product Quantity8
Reason for Recall
It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined that the incorrect insert was used which indicates that the device is MR Safe, when it is actually MR Conditional.
Firm Action
Synthes sent a Urgent Notice Medical Device Recall letter dated September 28, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
The firm is requesting that you please examine your inventory for products with the above Part and Lot number and remove them. Call Synthes at 1-800-479-6329.
If you have any further questions please call 610-719-5450.
Root Cause
Packaging
Distribution Pattern
US Distribution including the states of CA, MN, NC and TN.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.