Nordson Micromedics FibriJet Endoscopic Catheter Introducer, Catalog Number SA-3445. Sterilized using Ethylene Oxide
FDA Recall Z-1817-2017 · Posted None
Advertisement
Recalling FirmMicromedics, Inc.
Recall StatusTerminated
Product CodeKGZ
Date Initiated2016-10-20
Date Terminated2019-07-23
Product Quantity273 units
Reason for Recall
The sterile packaging may contain small channels in the seal of the pouch, If the channels are present, the package integrity may be compromised which may affect sterility assurance.
Firm Action
A Medical Device Recall Letter was sent to customers beginning 10/20/2016 via Federal Express in the US and via UPS ground in Israel. The letter identified the affected device and lot number, issue, asked for product to be discontinued from use, quarantine and notify Nordson Medical, via a response form. Response form can be faxed back at 888-504-0606, email to [email protected] or mailed to Nordson Medical, 1270 Eagan Industrial Road, Suite 120, St. Paul, MN 55121.
Root Cause
Employee error
Distribution Pattern
AL, CA, FL, MS, NC, NJ, NY, TN, and TX. Outside the US: Israel