da Vinci Xi EndoWrist Suction Irrigator, 8 mm instrument; PN 480299-03; and SOFTWARE, EMBEDDED RLS, IS4000, A70_P6_B440; PN 610092-440. General and Plastic Surgery: The EndoWrist Suction Irrigator is designed to be used in conjunction with an Intuitive Surgical da Vinci Surgical System and compatible suction and irrigation sources and tubing sets for delivering fluid to the surgical site and for evacuation and aspiration of fluids. The instrument may also be used for retraction and blunt dissection of tissue. The instrument tip is blunt and intended to contact tissue.
FDA Recall Z-1819-2017 · Posted 2017-04-13
Advertisement
Recalling FirmIntuitive Surgical, Inc.
Recall StatusTerminated
Product CodeNAY
Date Initiated2017-03-31
Date Terminated2017-07-28
Product Quantity15 x 6 packs (90 units)
Reason for Recall
Intuitive Surgical has become aware that in specific scenarios with system software P6 and the da Vinci Xi
Suction Irrigator (PN 480299-03), users can experience unexpected motion of a system arm.
Firm Action
Firm sent e-mails on 3-17-17 to affected sites instructing them to STOP USE of the XI Suction Irrigator and RMA. and return all unused Xi Suction/irrigation instruments for credit. A formal letters was sent to affected sites on March 31, 2017. Letters advised of the reason for recall with details on product name and product number. Letters requested that any unused devices be returned to Intuitive via the standard RMA process. All personnel who use the Da Vinci system should be made aware of the issue. Response form should be completed online or by returning copy enclosed with the letter.
Root Cause
Software design
Distribution Pattern
US Only - one location each in AL, CO, KS, NV, NY, and 2 in TX
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.