Brilliance 64 S/N 9089, 9551, 9548. The Brilliance 64 is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
FDA Recall Z-1820-2011 · Posted 2011-03-24
Advertisement
Recalling FirmPhilips Medical Systems (Cleveland) Inc
Recall StatusTerminated
Product CodeJAK
Date Initiated2010-12-20
Date Terminated2013-02-22
Product Quantity3
Reason for Recall
During the version 2.6 software upgrade, the system was not recalibrated with the infant phantom for the infant protocol.
Firm Action
Philips sent an URGENT - Medical Device Correction letter dated January 31, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers may consider reviewing any infant images that may have been affected during the Field Test to confirm reading accuracy.
For further information or support concerning this issue, customers could contact their local Philips representative or local Philips Healthcare office for North America and Canada contact the Customer Care Solutions Center (1-800-722-9377, option 5: Enter Site ID or follow the prompts)
Root Cause
Other