Dall Miles Cable Grip System; Catalog number 6704-0-210; Stryker Ireland. The Dall-Miles Recon and Trauma Cable System provides the surgeon with a variety of methods for achieving trochanteric reattachment and for cerclage fixation.

FDA Recall Z-1821-2009 · Posted 2009-08-18

Recall StatusTerminated
Product CodeKTT
Date Initiated2009-06-18
Date Terminated2012-08-27
Product Quantity13 total, (both products)

Reason for Recall

Sterilization issue: During post-sterilization inspection of the outer blister packaging of the affected lots, it was determined that the outer blister did not meet the required minimum seal width.

Firm Action

Urgent Product Recall letters were sent via Federal Express on June 18, 2009 to all Stryker Branches/Agencies, hospital Risk Management, hospital Chief of Orthopaedics and surgeons who may have used the recalled products. The letters stated the issue, potential hazards, and the risk mitigation. Customers may contact Colleen O'Meara at 201-831-5970 with any questions.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution