V 40 Femoral Head, Sterile, 22 mm; Catalog number 6260-4-122, Catalog number 6260-5-226; Howmedica Osteonics Corp., Stryker Ireland, Carrigtwohill County Cork, Ireland; Intended for use with femoral stems and acetabular components in primary or revision total hip arthroplasty.
FDA Recall Z-1822-2009 · Posted 2009-08-18
Advertisement
Recalling FirmStryker Howmedica Osteonics Corp.
Recall StatusTerminated
Product CodeLWJ
Date Initiated2009-06-18
Date Terminated2012-08-27
Product Quantity13 total, both products.
Reason for Recall
Sterilization issue: During post-sterilization inspection of the outer blister packaging of the affected lots, it was determined that the outer blister did not meet the required minimum seal width.
Firm Action
Urgent Product Recall letters were sent via Federal Express on June 18, 2009 to all Stryker Branches/Agencies, hospital Risk Management, hospital Chief of Orthopaedics and surgeons who may have used the recalled products. The letters stated the issue, potential hazards, and the risk mitigation. Customers may contact Colleen O'Meara at 201-831-5970 with any questions.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.