Terminated

The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010; Allura Xper FD10/10 722011; Allura Xper FD20 722012; Allura Xper FD20 biplane 722013; Allura Xper FD10 OR Table 722014; Allura Xper FD20 OR Table 722015; Allura Xper FD10/10 OR Table 722019; Allura Xper FD20 Biplane OR Table 722020; Allura Xper FD10 OR Table 722022; Allura Xper FD20 OR Table 722023; Allura Xper FD10/10 OR Table 722024; Allura Xper FD20 Biplane OR Table 722025; Allura CV20 722031

FDA Recall Z-1822-2017 · Posted None

Recall StatusTerminated
Product CodeOWB
Date Initiated2017-03-22
Date Terminated2021-05-17
Product Quantity6992 (total distributed for all products)

Reason for Recall

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Firm Action

Consignees were notified of the recall via letter sent on 3/22/2017. The letter explains the issue, actions that should be taken by the customer, and actions to be taken by Philips.

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide, worldwide including Canada