Terminated

Philips MR systems: Ingenia HA FLEX TRAK Trolley (for Ingenia 1.5T & 3.0T products) Product Usage: Diagnostic MR System. The Height Adjustable FlexTrak trolley and Trolley Variable Height IRF is an optional device designed to prepare the patient outside the examination room to enhance workflow and make additional non-magnetic hospital beds unnecessary.

FDA Recall Z-1826-2014 · Posted 2014-06-18

Recall StatusTerminated
Product CodeLNH
Date Initiated2014-05-15
Date Terminated2015-08-07
Product Quantity420

Reason for Recall

Ingenia customers have experienced clamping of the foot under the central column of the Height Adjustable Flex Trak trolley and the Trolley Variable Height IRF.

Firm Action

Philips sent an Urgent Field Safety Notice (FSN78100414) dated May 2014 via certified mail to US and Canadian consignees. Philips Key Market representatives will distribute the letter internationally. The letter identified the affected product, explains the problem, actions that should be taken in order to prevent risks for users, and the actions planned by Philips to correct the problem. On May 31, 2014, Philips released a Field Change Order that requires service engineers to replace the front and rear covers of the Ingenia MR FlexTrak Trolley with new covers with protection bumper that prevents foot clamping. For further information and support concerning this issue, consignees should contact their local Philips representative at 1-800-722-9377.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US (nationwide), Canada, Mexico Australia, Austria, Germany, Belarus, Bulgaria, Brazil, China, Denmark, Estonia, Finland, France, Germany, Greece, Italy, Jordan, Korea, Republic of Korea, Netherlands, New Zealand, Norway, Oman, Pakistan, Russian Federation, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.